Medical device recalls Moderate risk

WalkMed Infusion, LLC recalls Triton Infusion Pump (model 300000) and Triton fp Infusion Pump (model 400000) (proprietary name: AFF (Mark I) Volumetr…

Recall date
October 1, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0242-2016
FDA classification
Class II
Brand / firm
WalkMed Infusion, LLC
Sold / distributed
Distributed in the states of AL, CA, CO, DE, FL, GA, IL, MA, MD, MI, MN, MS, NC, NJ, NM, NV, NY, OH, PA, TN, TX, and WA.

Why it was recalled

The device might fail to detect air in line while infusing the medication Venofer. This may occur when a drop of Venofer gets stuck in the tubing adjacent to the bubble detector and the Total Volume Limit is programmed higher than the actual bag volume.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Triton Infusion Pump (model 300000) and Triton fp Infusion Pump (model 400000) (proprietary name: AFF (Mark I) Volumetric Infusion Pump). Used to pump fluids into a patient in a controlled manner.

Get recall alerts

Free email alert whenever WalkMed Infusion, LLC has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: WalkMed Infusion, LLC