WELCH ALLYN, INC/MORTARA recalls ELI PC. The ELI PC Service automatically converts resting ECG recordings collected with WAM PC patient cables and WebUp…
- Recall date
- March 18, 2020
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1811-2020
- FDA classification
- Class II
- Brand / firm
- WELCH ALLYN, INC/MORTARA
- Sold / distributed
- US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.
Why it was recalled
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
ELI PC. The ELI PC Service automatically converts resting ECG recordings collected with WAM PC patient cables and WebUpload into DICOM 12-lead ECG Waveform files.