Medical device recalls Moderate risk

WELCH ALLYN, INC/MORTARA recalls Lead diagnostic electrocardiograph under the following device names: ELI 10, ELI 350, ELI 150c, ELI 250c, ELI 230, ELI…

Recall date
March 18, 2020
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1810-2020
FDA classification
Class II
Brand / firm
WELCH ALLYN, INC/MORTARA
Sold / distributed
US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.

Why it was recalled

Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Lead diagnostic electrocardiograph under the following device names: ELI 10, ELI 350, ELI 150c, ELI 250c, ELI 230, ELI 280, ELI 380, R-Scribe and Connex Cardio. Product Usage: A non-invasive prescription device indicated for use to acquire, analyze, display, transmit and print electrocardiograms.