Medical device recalls Moderate risk

WELCH ALLYN, INC/MORTARA recalls S4 and patient cable for the S4 Wireless Telemeter with the following device names: CABLE 10 WIRE LF ECG SNAP ENDS AHA…

Recall date
March 18, 2020
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1815-2020
FDA classification
Class II
Brand / firm
WELCH ALLYN, INC/MORTARA
Sold / distributed
US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.

Why it was recalled

Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

S4 and patient cable for the S4 Wireless Telemeter with the following device names: CABLE 10 WIRE LF ECG SNAP ENDS AHA GRAY, CABLE 10 WIRE LF ECG SNAP ENDS IEC GRAY, CBL 10WIRE LF LRG ECG SNAP ENDS AHA GRAY, CBL 10WIRE LF LRG ECG SNAP ENDS IEC GRAY. The S4 mobile monitor provides a means to acquire and transmit simultaneous ECG and SpO2 data to a Surveyor Central Station monitoring system while allowing the patient to be ambulatory within the range of the WiFi network.