Medical device recalls Moderate risk

WELCH ALLYN, INC/MORTARA recalls Surveyor S12/S19 and patient cable for the S12/S19 Bedside Monitor under the following device names: PAT CBL 10WIRE AHA…

Recall date
March 18, 2020
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1816-2020
FDA classification
Class II
Brand / firm
WELCH ALLYN, INC/MORTARA
Sold / distributed
US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.

Why it was recalled

Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Surveyor S12/S19 and patient cable for the S12/S19 Bedside Monitor under the following device names: PAT CBL 10WIRE AHA SNAP JSCREW, PAT CBL 10WIRE IEC SNAP JSCREW. The Surveyor S12 patient monitor is designed for continuous monitoring in either stationary or portable applications.