WELCH ALLYN, INC/MORTARA recalls X12+ and patient cable for the X12+ Telemeter with the following device name: PAT CBL 10WIRE LEADFORM AHA SNAP, PAT CBL…
- Recall date
- March 18, 2020
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1814-2020
- FDA classification
- Class II
- Brand / firm
- WELCH ALLYN, INC/MORTARA
- Sold / distributed
- US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.
Why it was recalled
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
X12+ and patient cable for the X12+ Telemeter with the following device name: PAT CBL 10WIRE LEADFORM AHA SNAP, PAT CBL 10WIRE LEADFORM IEC SNAP, PAT CBL 10WIRE LEADFORM XL AHA SNAP, PAT CBL 10WIRE LEADFORM XL IEC SNAP, PAT CBL 4WIRE AHA SNAP HC.