Wipro GE Healthcare Private Ltd. recalls GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5; Part Number 5867474-04;…
- Recall date
- September 5, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0213-2026
- FDA classification
- Class II
- Brand / firm
- Wipro GE Healthcare Private Ltd.
- Sold / distributed
- Worldwide distribution.
Why it was recalled
GE HealthCare has identified an issue affecting Carescape" Central Station (CSCS) with software version V3.0.5, which can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5; Part Number 5867474-04; Medical Display, cathode ray tube