Medical device recalls Moderate risk

Wright Medical Technology, Inc. recalls Wright EVOLVE TRIAD Plate Cutter, REF 49510120

Recall date
June 14, 2022
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1366-2022
FDA classification
Class II
Brand / firm
Wright Medical Technology, Inc.
Sold / distributed
Nationwide and PR, and Australia, Canada, Chile, Colombia, Hong Kong, Puerto Rico

Why it was recalled

The plate cutter is unable to meet sterilization requirements when it is sterilized in a double-wrapped configuration according to the Option 1 Steam Sterilization parameters detailed in instructions for use

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Wright EVOLVE TRIAD Plate Cutter, REF 49510120