Xcision Medical Systems, LLC recalls GammaPod Stereotactic Radiotherapy System Model A, REF XMSGP030A00-0.02 Product Usage: GammaPod is a teletherapy device…
- Recall date
- July 3, 2019
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2081-2019
- FDA classification
- Class II
- Brand / firm
- Xcision Medical Systems, LLC
- Sold / distributed
- The products were distributed to the following US states: MD and TX.
Why it was recalled
One bolt on the V motor was not fully tightened which caused a drift of .13 degrees in alignment resulting in reduced dose output.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
GammaPod Stereotactic Radiotherapy System Model A, REF XMSGP030A00-0.02 Product Usage: GammaPod is a teletherapy device intended for use in the noninvasive stereotactic delivery of a radiation dose to a partial volume of the breast in conjunction with breast conserving treatment.
Get recall alerts
Free email alert whenever Xcision Medical Systems, LLC has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Xcision Medical Systems, LLC