Medical device recalls

Xintec Corp recalls OptiLITE, Products for Laser Surgery, Model # CGH1080F, 800 micron single-use holmium fiber assembly with flat tip, Lot…

Recall date
October 29, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0162-2020
FDA classification
Class III
Brand / firm
Xintec Corp
Sold / distributed
PA and VA

Why it was recalled

It has been determined that a lot of fiber lasers were programmed incorrectly. As a result, the fibers from this lot will produce an error and will not allow the user to use the laser when it is plugged in.

Class III means the product violates regulations but is not likely to cause harm.

What was recalled

OptiLITE, Products for Laser Surgery, Model # CGH1080F, 800 micron single-use holmium fiber assembly with flat tip, Lot Number 21144, Sterilized by Sterrad Run 4149, This device is for single-use only, sterile, UDI: 00855430006036