Zap Surgical Systems recalls ZAP-X Radiosurgery System, Model: 300150
- Recall date
- July 25, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2639-2025
- FDA classification
- Class II
- Brand / firm
- Zap Surgical Systems
- Sold / distributed
- Worldwide - US Nationwide distribution in the states of FL, CO, NJ, NY and the countries of Germany, France, Spain, Poland, Republic of Korea, T¿rkiye, Paraguay.
Why it was recalled
If the radiosurgery system triggers a proximity error message during a long gantry move (greater-than180 degrees), and a subsequent proximity error message occurs after the automatic reduction of speed, a software defect could potentially lead the collimator to collide with patient shoulders or the patient table.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
ZAP-X Radiosurgery System, Model: 300150
Get recall alerts
Free email alert whenever Zap Surgical Systems has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Zap Surgical Systems