Medical device recalls Moderate risk

Zeiss, Carl Inc recalls Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope

Recall date
May 29, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2441-2018
FDA classification
Class II
Brand / firm
Zeiss, Carl Inc
Sold / distributed
US Nationwide Distribution

Why it was recalled

Under certain circumstances, the firmware makes it possible for the Disinfection unit UV (432332-9020-000) to activate outside of the Celldiscoverer 7 housing. This may result in exposure of the users to harmful UV radiation.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope