Ziehm Imaging Inc recalls Ziehm Vision RFD 3D Mobile C-arm for Mobile Fluoroscopic Imaging
- Recall date
- April 15, 2016
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2639-2016
- FDA classification
- Class II
- Brand / firm
- Ziehm Imaging Inc
- Sold / distributed
- US Distribution including Puerto Rico and to the states of :FL, GA, VA,MA. ME, MI, NY, NJ, PA, RI, IL, TX, AZ, CA, NV, UT, and WA.
Why it was recalled
Malfunction of the radiation-warning timer and the 3D-specific interval warning. No 5-minute timer dialogs or beeper will occur until the unit is rebooted.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Ziehm Vision RFD 3D Mobile C-arm for Mobile Fluoroscopic Imaging