Medical device recalls Moderate risk

Zien Medical Technologies, Inc recalls PenBlade Safety Scalpel, for the incision of human skin and tissue, and to cut or trim sutures and catheters up to 6Fr.…

Recall date
August 18, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2800-2015
FDA classification
Class II
Brand / firm
Zien Medical Technologies, Inc
Sold / distributed
Worldwide distribution to U.S. nationwide, Germany and England.

Why it was recalled

Cracks were discovered in the PET tray of a small population of sterile product. These cracks create a risk of loss of sterile barrier for the product. Sampling and root cause analysis indicated that the risk could extend throughout the lots manufactured from September 2014-July 2015. This led to the decision to recall the product.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

PenBlade Safety Scalpel, for the incision of human skin and tissue, and to cut or trim sutures and catheters up to 6Fr. Models PB-M-10, PB-M-11P, PB-M-15

Get recall alerts

Free email alert whenever Zien Medical Technologies, Inc has a new recall — straight from official government data. Unsubscribe anytime.