Zien Medical Technologies, Inc recalls PenBlade Safety Scalpel, for the incision of human skin and tissue, and to cut or trim sutures and catheters up to 6Fr.…
- Recall date
- August 18, 2015
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2800-2015
- FDA classification
- Class II
- Brand / firm
- Zien Medical Technologies, Inc
- Sold / distributed
- Worldwide distribution to U.S. nationwide, Germany and England.
Why it was recalled
Cracks were discovered in the PET tray of a small population of sterile product. These cracks create a risk of loss of sterile barrier for the product. Sampling and root cause analysis indicated that the risk could extend throughout the lots manufactured from September 2014-July 2015. This led to the decision to recall the product.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
PenBlade Safety Scalpel, for the incision of human skin and tissue, and to cut or trim sutures and catheters up to 6Fr. Models PB-M-10, PB-M-11P, PB-M-15