Medical device recalls Moderate risk

Zimmer Biomet, Inc. recalls Fem IM Nail 15mmdx50cm, Sterile, Rod, fixation, intramedullary and accessories- Specific Knee, Hip and Nail Implants

Recall date
February 28, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2559-2017
FDA classification
Class II
Brand / firm
Zimmer Biomet, Inc.
Sold / distributed
Worldwide Distribution - US (nationwide) Foreign: Argentina, Australia, Belgium, Brazil, Bridgend Wales, Canada, Chile, Colombia, Dominican Republic, England, Eschbach, France, India, Italy, Japan, Jordan, Korea, Malaysia, Mexico, Netherlands, Nicaragua, Panama, Peru, Russia, Saudi Arabia, Singapor…

Why it was recalled

Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Fem IM Nail 15mmdx50cm, Sterile, Rod, fixation, intramedullary and accessories- Specific Knee, Hip and Nail Implants