Medical device recalls Moderate risk

Zimmer Biomet, Inc. recalls NCB-PH Plate with 4/5/7 Lock Holes, Item Nos. 0202262104 0202262105 0202262107

Recall date
October 10, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0332-2020
FDA classification
Class II
Brand / firm
Zimmer Biomet, Inc.
Sold / distributed
Distributed nationwide.

Why it was recalled

Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration).

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

NCB-PH Plate with 4/5/7 Lock Holes, Item Nos. 0202262104 0202262105 0202262107