Medical device recalls Moderate risk

Zimmer Biomet, Inc. recalls Uniflex Humeral Nail: 467265 G-PIN 2.8BL 3.2PL 60CM 2PK Pin, Fixation, Smooth A 467265-01 Ball Tip Guide Pin is used to…

Recall date
August 2, 2016
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0087-2017
FDA classification
Class II
Brand / firm
Zimmer Biomet, Inc.
Sold / distributed
US Distribution to states of: MD,TX, FL, GA, VA and PA.

Why it was recalled

The G-PIN 2.8BL 3.2PL 60 CM 2PK is intended to contain one (1) 467265-01 Ball Tip Guide Pin and one (1) 467255-02 Plain Tip Guide Pin. The G-PIN 2.8BL 3.2PL 60 CM 2PK may have been packaged incorrectly either with two (2) 467265-01 Ball Tip Guide Pins and zero (0) 467255-02 Plain Tip Guide Pins or with zero (0) 467265-01 Ball Tip Guide Pins and two (2) 467255-02 Plain Tip Guide Pins. If the G-PIN 2.8BL 3.2PL 60 CM 2PK contains the incorrect combination of Plain Tip Guide Pins and Ball Tip Guide Pins, the surgeon would either have to implant the nail without the Plain Tip Guide Pin as a guide, ream with a Plain Tip Guide Pin, or choose to not ream the humeral canal. The risk associated with not having the correct combination can include a delay in surgery of less than 30 minutes, the reamer tip remaining in the distal humeral canal, or an iatrogenic fracture.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Uniflex Humeral Nail: 467265 G-PIN 2.8BL 3.2PL 60CM 2PK Pin, Fixation, Smooth A 467265-01 Ball Tip Guide Pin is used to guide reamers down the humeral canal during a Uniflex Humeral Nail procedure. A 467255-02 Plain Guide Pin is used to guide a Uniflex Humeral Nail implant into the humeral canal after the fracture has been reduced.

Get recall alerts

Free email alert whenever Zimmer Biomet, Inc. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Zimmer Biomet, Inc.