Medical device recalls Moderate risk

Zimmer Biomet, Inc. recalls Vital MIS Spinal Fixation system implants CANN POLY EXT TAB 6.5X40MM, Item Number 810M6540 - Product Usage: are non-cer…

Recall date
May 5, 2020
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2224-2020
FDA classification
Class II
Brand / firm
Zimmer Biomet, Inc.
Sold / distributed
US nationwide distribution.

Why it was recalled

There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Vital MIS Spinal Fixation system implants CANN POLY EXT TAB 6.5X40MM, Item Number 810M6540 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle screw fixation, posterior hook fixation, or anterolateral fixation. Pedicle screw fixation is indicated for skeletally mature patients and for adolescence patients.

Get recall alerts

Free email alert whenever Zimmer Biomet, Inc. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Zimmer Biomet, Inc.