Medical device recalls Moderate risk

Zimmer Biomet, Inc. recalls Zimmer, Virage OCT Spinal Fixation System, Item Number: 07.01738.001, a posterior system intended to achieve an implant…

Recall date
November 17, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0625-2016
FDA classification
Class II
Brand / firm
Zimmer Biomet, Inc.
Sold / distributed
US: AL, AZ, CA, CO, FL, GA, ID, MI, MN, MO, NV, NJ, OK, OR, PA, SD, TN, TX, WA, WI. OUS: AUSTRALIA, GERMANY, SWITZERLAND, SWEDEN.

Why it was recalled

Zimmer Biomet is initiating a recall of the Virage Occipital Eyelet, as it was found through product testing that the staking of the set screw was out of specification and can interfere with the insertion of the occipital rod when the set screw is staked too low.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Zimmer, Virage OCT Spinal Fixation System, Item Number: 07.01738.001, a posterior system intended to achieve an implant and construct of the Occipital-Cervical-Thoracic spine (Occiput-T3). UDI Number 889024328952

Get recall alerts

Free email alert whenever Zimmer Biomet, Inc. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Zimmer Biomet, Inc.