Zyno Medical LLC recalls Zyno Medical Administration Set -105"ADMIN SET 20DP, W/2 Y SITE, CVCLMPS- For Use with Zyno Medical Infusion pumps or G…
- Recall date
- May 29, 2019
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2058-2019
- FDA classification
- Class II
- Brand / firm
- Zyno Medical LLC
- Sold / distributed
- MA
Why it was recalled
Administration sets assembled with inverted pinch clamp cause the fluid from the distal end of the administration set to be pumped to the proximal end of the tubing when secondary set is used with the primary set. Thus, instead of delivering fluid to the patient, blood may be drawn from the patient if there is space in the drug container to hold extra volume of fluid, under the condition that the reversed pinch clamp is not identified
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Zyno Medical Administration Set -105"ADMIN SET 20DP, W/2 Y SITE, CVCLMPS- For Use with Zyno Medical Infusion pumps or Gravity Feed Product Code:B2-70072
Get recall alerts
Free email alert whenever Zyno Medical LLC has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Zyno Medical LLC