Medical device recalls Moderate risk

Zyno Medical LLC recalls Zyno Medical Administration Set -105"ADMIN SET 20DP W/FILTER 2YSITES CLMP- For Use with Zyno Medical Infusion pumps or…

Recall date
May 29, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2059-2019
FDA classification
Class II
Brand / firm
Zyno Medical LLC
Sold / distributed
MA

Why it was recalled

Administration sets assembled with inverted pinch clamp cause the fluid from the distal end of the administration set to be pumped to the proximal end of the tubing when secondary set is used with the primary set. Thus, instead of delivering fluid to the patient, blood may be drawn from the patient if there is space in the drug container to hold extra volume of fluid, under the condition that the reversed pinch clamp is not identified

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Zyno Medical Administration Set -105"ADMIN SET 20DP W/FILTER 2YSITES CLMP- For Use with Zyno Medical Infusion pumps or Gravity Feed Product Code: B2-70072-F