SIGHT DIAGNOSTICS LTD recalls

4 recalls on record · latest: May 13, 2024

Official U.S. recall history for SIGHT DIAGNOSTICS LTD, compiled from official government records.

Get recall alerts

Free email alert whenever SIGHT DIAGNOSTICS LTD has a new recall — straight from official government data. Unsubscribe anytime.

SIGHT DIAGNOSTICS LTD recalls Sight OLO CBC Test Kit, Model/Catalog Number: TK1. A disposable cartridge and sample preparation tools for use with the…

May 13, 2024 · Medical device recalls Moderate risk One Test Kit lot (TK1-10114) shows a positive bias in the platelet count due to…

SIGHT DIAGNOSTICS LTD recalls Sight OLO, Automated Hematology Analyzer, Catalog No. OLO-U1

June 14, 2022 · Medical device recalls Moderate risk The default reference ranges for MCH, MCHC, and RDW for age group "Child 2-11y"…

SIGHT DIAGNOSTICS LTD recalls The Sight OLO device is a computer vision based platform for blood analysis. The platform combines computer-vision algo…

January 20, 2022 · Medical device recalls Moderate risk In instances where custom reference ranges were configured on the device post i…

SIGHT DIAGNOSTICS LTD recalls Sight OLO-E1/U1 Software and Operators Manual Software v2.56.1 Catalog #: 1 OLO-U1 Catalog #: 2 OLO-E1 revision 1.9, re…

March 17, 2021 · Medical device recalls Moderate risk Discrepancy in labeling between the adult CBC reference ranges presented in the…