Medical device recalls Moderate risk

SIGHT DIAGNOSTICS LTD recalls Sight OLO CBC Test Kit, Model/Catalog Number: TK1. A disposable cartridge and sample preparation tools for use with the…

Recall date
May 13, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2173-2024
FDA classification
Class II
Brand / firm
SIGHT DIAGNOSTICS LTD
Sold / distributed
Worldwide - US Nationwide distribution in the states of AL, FL, MS, TX, WI and the countries of Canada, Israel, Slovenia, Trinidad & Tobago.

Why it was recalled

One Test Kit lot (TK1-10114) shows a positive bias in the platelet count due to bacterial contamination which can result in elevated counts with a bias of 7.5 k/ul. This results in the test kit lot performing outside of the device specification.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Sight OLO CBC Test Kit, Model/Catalog Number: TK1. A disposable cartridge and sample preparation tools for use with the Sight OLO CBC analyzer.

Get recall alerts

Free email alert whenever SIGHT DIAGNOSTICS LTD has a new recall — straight from official government data. Unsubscribe anytime.